Technical specifications
2001989-351A ApexPro™ A-11
Environmental specifications
Operating conditions
Transport and storage conditions
Physical specifications
Certification
UL60601 Medical Equipment with respect to electric shock, fire and mechanical
hazards only in accordance with UL 60601-1, and CAN/CSA C22.2 No. 601.1, and
IEC 60601-1
Start key: Allows the patient or caregiver to initiate readings on
demand.
Power source: Four 1.5 (AA) alkaline batteries. Rechargeable batteries are
not recommended (sample quantity with rechargeable
batteries typically limited to 25 samples/measurements vs.
250 samples/measurements with new alkaline batteries).
Temperature: +10°C to 40°C (50°F to 104°F)
Relative Humidity: <95%
Temperature: -20°C to +50°C (-4°F to +122°F)
Relative Humidity: <95%
Power: 6 Vdc, 4 1.5V AA Alkaline batteries.
Length: 12.7 cm (5.0 in)
Width: 8.25 cm (3.25 in)
Thickness: 3.3 cm (1.30 in)
Weight: 357.2 g (12.6 oz)