Technical specifications
2001989-351A ApexPro™ A-3
Environmental specifications
Operating conditions
Transport and storage conditions
Device specifications
QRS detection range: 0.5 to 5 mV
Frequency response: up to 40 Hz (–3dB)
Ambient temperature: 0°C to 50°C (32°F to 122°F)
Relative humidity: 15% to 95% (non-condensing)
Temperature: –40°C to 70°C (-40°F to 158°F)
Relative humidity: 5% to 95% (non-condensing)
Pressure: 475 to 1200 hPa
Water resistance: ApexPro model: IEC 60529 IPX3 rating (spray and
wipe only)
ApexPro CH model: IEC 60529 IPX7 rating (can
survive inadvertent submersion)
Frequency range: ApexPro model:
(The frequency range depends on the PCB installed.)
420–460 MHz or
584–613.975 MHz (programmable synthesizer)
ApexPro CH model:
608.025–613.975 MHz
Power output: ApexPro model:
(Meets FCC requirements for Parts 15 and 95.)
0.64mW @ 420–460
0.5mW @ 584-614 MHz
ApexPro CH model:
(Meets FCC requirements for Part 95.)
0.5mW @ 608.025–613.975 MHz
Bandwidth ApexPro model:
9.5 KHz
ApexPro CH model:
9.7 KHz
Modulation: ApexPro model: GMSK
ApexPro CH model: GFSK
Alarm Pause: Transmitted and indicated via LED

Table of Contents

Other manuals for GE ApexPro

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ApexPro and is the answer not in the manual?

GE ApexPro Specifications

General IconGeneral
BrandGE
ModelApexPro
CategoryMedical Equipment
LanguageEnglish

Related product manuals