GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 1 Direction 18000
6-3-5
3-3 Mag Tape Adjustments (Continued)
E. EOT/BOT Sensor Alignment
Refer to the procedure in the 6026 Manual, Part 4, Chapter V.
F. Head Alignment
Refer to the procedure in the 6026 Manual, Part 4-Maintenance Manual, Chapter V. When necessary load
the Command String Interpreter as described in procedure A.
To read the skew tape, type the underlined responses:
UNIT 0 ’CR’
TAPE MODE NRZI ’CR’
WC 37777 ’CR’
DATA ’CR’
COMMAND STRING
REA 500 SPA BAC 500 ’CR’
To return the skew tape to BOT at 75 ips, not using rewind:
UNIT 0 ’CR’
TAPE MODE ’CR’
WC 37777 ‘CR’
DATA ‘CR’
COMMAND STRING
SPA BAC 1000 ‘CR’
Repeat, if needed, to reach BOT.
To write all 1s on a blank tape:
UNIT 0 ’CR’
TAPE MODE NRZI ’CR’
WC 37777 ’CR’
DATA ALL ONE ’CR’
COMMAND STRING
REW WRI 500 ’CR’