GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 1 Direction 18000
6-3-6
3-3 Mag Tape Adjustments (Continued)
G. Read Gain Master Level Adjustment
Refer to the 6026 Manual, Part 4, Chapter V for details of the procedure. When necessary, the command
string may be loaded as described in Procedure A.
To write all 1s, type the underlined:
UNIT 0 ’CR’
TAPE MODE NRZI ’CR’ or
PE ’CR’ (as needed)
WC 37777 ’CR’
DATA ALL ONE ’CR’
COMMAND STRING
REW WRI 100 ’CR’
H. NRZI Read Data Recovery Adjustment
Refer to the 6026 Manual, Part 4, Chapter V for details of the procedure. When necessary, the command
string may be loaded as described in Procedure A.
To write all 1s, type the underlined:
UNIT 0 ’CR’
TAPE MODE NRZI ’CR’
WC 37777 ’CR’
DATA ALL ONE ’CR’
COMMAND STRING
REW WRI 100 ’CR’
I. Reel Servo (Vacuum Level) Adjustment
Refer to the procedure in the 6026 Manual, Part 4, Chapter V.