GE MEDICAL SYSTEMS CT 9800 QUICK SYSTEM
Rev. 3 Direction 18000
8-3-1
SECTION 3 ERROR LOG
3-1 Entering the Log
A log of all system errors is available for review. At R Level type M/S return and press the “MANAGE” key (M/S
is for Service Use Only,) not for Customer use. Customers should use M level only. (M/S allows option 6 to
appear on the menu,) and respond as follows:
______________________________________
MANAGE MENU
ENTER 6 return
SPECIFY SELECTION =>_6_
1 = REMOVE
2 = TUBE WARM-UP
3 = CALIBRATION
4 = SETUP FILES
5 = SYSTEM STATUS
6 = SERVICE SUPPORT
______________________________________
MANAGE MENU
SERVICE SUPPORT
ENTER 1 return
SPECIFY SELECTION =>_1_
1 = EXAMINE ERROR LOG
2 = TERMINATE/START ERROR LOG
3 = N/A
4 = N/A
5 = EXAMINE DEVICE STATUS
6 = EXAMINE DEVICE STATUS
_____________________________________
EXAMINE ERROR LOG
CURRENT LOG? (Y/N) => _Y_
ENTER Y return