1-60 Programming Site Data
Entering KV vs. SS Tables The KV/SS submenu enables you to enter a table of values of the
kinematic viscosity (KV) based on signal strength (SS). You can then
use this table to calculate KV as shown in the KV/SS option (in the
ADVAN option of the SETUP submenu) discussed on page 1-33. To use
the option, you must enter the KV vs. SS pairs in descending order of
KV and ascending order of SS. For example, if pair 1 is KV = 10
centistrokes and SS = 62, then pair 2 must be KV ≤ 10 centistrokes
and SS 62. The DF868 accepts 2 to 20 pairs.
1. To enter the
GLOBAL menu, press [F3]=GLOBL.
1. To enter the KV/SS submenu, press [→] and [F1].
2. Press
[ENT] to accept the current # of KV/SS Pairs or enter a new
value (from 2 to 20) and press [ENT].
3. Press
[ENT] to accept the current Sig Strength value or enter a new
value (50.0 to 85.0) and press
[ENT]. Enter the signal strength
values in ascending order.
4. Press
[ENT] to accept the current Kin Viscosity value or enter a new
value (0.050 to 500.) and press
[ENT]. Enter the kinematic viscosity
values in descending order.
The Sig Strength and Kin Viscosity prompts repeat for each pair. After
entering all the pairs, the meter returns to the
Global PROGRAM menu.

Table of Contents

Other manuals for GE DigitalFlow DF868

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE DigitalFlow DF868 and is the answer not in the manual?

GE DigitalFlow DF868 Specifications

General IconGeneral
BrandGE
ModelDigitalFlow DF868
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals