101
Page 101
Mini-AT User Guide
Item
No. Item Name Description
————————————————————————————————————————
099 BATTERY LOW This item (099) indicates if a low voltage alarm for the main
battery was generated. During a wake-cycle, if the
measurement for item 048 (BATTERY VOLTAGE) is a value
less than the value at item 049 (BATTERY LOLIM V), an
alarm is initiated and is indicated by placing dots on the LCD
and “11111111” at item 099. “00000000” at item 099
indicates there is no Low Battery Alarm. The alarm indicators
will remain active until manually cleared (by software) or
automatically when RBX (item 165) is enabled.
Default = 00000000
100 BATT CYCLES ALRM This item (100) indicates if a high wake-cycle alarm for the
main battery was generated. During a wake-cycle, if the
calculated value for item 059 (BATTERY WAKE-UP
CYCLES) is a value greater than the value at item 060 (BATT
CYCLES LIM), an alarm is initiated and is indicated by
placing dots on the LCD and “11111111” at item 100.
“00000000” at item 100 indicates there is no Battery Cycles
Alarm. The alarm indicators will remain active until manually
cleared (by software).
Default = 00000000
101 BKUP BATTERY LOW This item (101) indicates if a low voltage alarm for the backup
battery was generated. During a wake-cycle, if the
measurement for item 051 (MEMORY BATTERY
VOLTAGE) is a value less than the value at item 052 (MEM
BATT LOVOLT LIMIT), an alarm is initiated and is indicated
by placing dots on the LCD and “11111111” at item 101.
“00000000” at item 101 indicates there is no Low Mem
Battery Alarm. The alarm indicators will remain active until
manually cleared (by software) or automatically when RBX
(item 165) is enabled.
Default = 00000000
102 INDEX SW#1 FAULT This item (102) indicates if an alarm for index switch #1 was
generated. If index switch #1 fails to provide an input meter
pulse within four meter revolutions, an alarm is initiated and is
indicated by placing dots on the LCDand “11111111” at item
101. “00000000” at item 101indicates there is no Switch #1
Alarm. The alarm indicators will remain active until manually
cleared (by software).
Default = 00000000