117
Page 117
Mini-AT User Guide
Item
No. Item Name Description
————————————————————————————————————————
190 DAILY ENERGY Same as item 140 (Energy), except this value is the running total
for the current Gas Day, which is zeroed and begins a new
calculation at the time entered at item 205. Therefore, the value
displayed during instrument access is the most recent calculated
value for the current gas day.
Default = 00000000
191 PREV DAY ENERGY Yesterday’s Daily Energy, i.e., the daily energy that is one ‘Gas
Day’ old at the time the instrument was accessed. This information
is updated with the value from item 190 at the beginning of every
new gas day. Note: The gas day begins at the time entered at item
205.
Default = 00000000
192 DAILY PK FLOW RT Same as item 210 (Peak Flow Rate), except this value is the peak
flow rate for the current Gas Day, which is zeroed at the time
entered at item 205. Therefore, the value displayed during
instrument access is the peak flow rate, up to that point in time,
during the current gas day.
Default = 00000000
193 DAILY PK FLOW TM The ending-hour time the Daily Peak Flow Rate (192) occurred
during the current gas day.
Default = 00 00 00
194 PREDAY PK FLOW RT Yesterday’s Daily Peak Flow Rate, i.e., the daily peak flow rate
that is one ‘Gas Day’ old at the time the instrument was accessed.
This information is updated with the value from item 192 at the
beginning of every new gas day. Note: The gas day begins at the
time entered at item 205.
Default = 00000000
195 PREDAY PK FLO TM The ending-hour time the Previous Daily Peak Flow Rate (194)
occurred during the previous gas day.
Default = 00 00 00
196 EVENT USER ID An identification number inserted into the Event Log to identify
the person that caused the change(s) to the instrument’s
configuration. The default value is “0” but may be voluntarily
changed to any number between “0” & “99” (inclusive) when
prompted during instrument access.
Default = 0