EasyManuals Logo
Home>ITech>Measuring Instruments>IT8200 Series

ITech IT8200 Series Programming Guide

ITech IT8200 Series
137 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #121 background imageLoading...
Page #121 background image
WAVEform Subsystem
Copyright © ITECH Electronic Co., Ltd. 101
WAVEform:EDIT:THD:FORMula <formula>
This command is used to set and query the distortion factor calculation method of
the edited THD wave.
Syntax:
WAVEform:EDIT:THD:FORMula <formula>
Arguments
<%F|%R>
Query syntax:
WAVEform:EDIT:THD:FORMula?
Returns:
<CPD>
Example
WAVE:EDIT:THD:FORM %F
WAVE:EDIT:THD:FORM?
WAVEform:EDIT:THD:DATA <order>,<string>
This command is used to set and query the THD value and phase parameter of a
certain order of harmonics.
Syntax:
WAVEform:EDIT:THD:DATA <order>,<string>
Arguments
<2-50>,<"thd,phase">
Query syntax:
WAVEform:EDIT:THD:DATA? <order>
Returns:
<string>
Example
WAVE:EDIT:THD:DATA 2, "9.8,0.0"
WAVE:EDIT:THD:DATA? 2
WAVEform:EDIT:THD:SAVE:LOCal <filename>
This command is used to save the presently edited THD wave data to a local file,
and specify the file name (if the file name is the same, it is regarded as modifying
the present waveform, if the file name does not exist, it is regarded as creating a
new file)

Table of Contents

Other manuals for ITech IT8200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8200 Series and is the answer not in the manual?

ITech IT8200 Series Specifications

General IconGeneral
BrandITech
ModelIT8200 Series
CategoryMeasuring Instruments
LanguageEnglish