SYSTem Commands
Copyright © ITECH Electronic Co., Ltd. 5
Example
SYST:PRES
SYSTem:POSetup <CPD>
This command is used to set and query some parameters or working status when
the instrument is powered on.
RST: The default value indicates that the factory initialization value is displayed
when the instrument is powered on. The specific parameters are described in
*RST.
LAST_ON: Indicates when powered on, the instrument will remain the same
parameter settings as last time you turned off the instrument.
LAST_OFF: Indicates when powered on, the instrument will remain the same
settings as last time you turned off the instrument, but the output state is Off.
Syntax
SYSTem:POSetup <CPD>
Arguments
RST|LAST|LAST_OFF
Default value
RST
Returns
None
Example
SYST:POS RST
Query syntax
SYSTem:POSetup?
SYSTem:CLEar
This command is used to clear the error queue.
Syntax
SYSTem:CLEar
Example
SYST:CLE