Input Subsystem
Copyright © ITECH Electronic Co., Ltd. 73
Chapter16 Input Subsystem
INPut:COUPling <CPD>
This command is used to switch AC and DC mode.
Syntax:
INPut|OUTPut:COUPling <CPD>
Arguments
DC|AC
Query syntax:
INPut:COUPling?
Returns:
DC|AC
Example
INPut:COUPling AC
INPut:PHASe:LOSS <STATE>
This command is used to enable the phase loss state.
Syntax:
INPut:PHASe:LOSS <STATE>
Arguments
0|OFF|1|ON
Query syntax:
INPut:PHASe:LOSS?
Example
INPut:PHASe:LOSS 1
INPut:LINE:CONNection <CPD>
This command is used to set the connection status of the output. WYE is the star
connection method and DELTa is the triangle connection method.
Syntax:
INPut:LINE:CONNection <CPD>