SWEep Subsystem
Copyright © ITECH Electronic Co., Ltd. 106
Query syntax:
SWEep:LEVel:STARt? [A|B|C][MAXimum|MINimum]
Returns:
<NR2>
Example
SWEep:LEVel:STARt 100 //Set the starting value of sweep to 100.
SWEep:LEVel:STARt?
SWEep:LEVel:STOP <NRf+>[,<NRf+>,<NRf+>]
This command is used to set and query the stop value of sweep.
<NRf+>: stop setting value, or stop MINimum or MAXimum or DEFault setting.
Note:
In single-phase (including AC, ACDC, DC, DCAC), there is only one parameter;
in three-phase, three parameters must be taken.
Syntax:
SWEep:LEVel:STOP <NRf+>[,<NRf+>,<NRf+>]
Arguments
<NRf+>[,<NRf+>,<NRf+>]
Query syntax:
SWEep:LEVel:STOP? [A|B|C][MAXimum|MINimum]
Returns:
<NR2>
Example
SWEep:LEVel:STOP 90 // Set the stopping value of sweep to 100.
SWEep:LEVel:STEP <NRf+>[,<NRf+>,<NRf+>]
This command is used to set and query the step level of sweep.
<NRf+>: step setting value, or step MINimum or MAXimum or DEFault setting.
Note:
In single-phase (including AC, ACDC, DC, DCAC), there is only one parameter;
in three-phase, three parameters must be taken.
Syntax:
SWEep:LEVel:STEP <NRf+>[,<NRf+>,<NRf+>]

Table of Contents

Other manuals for ITech IT8200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8200 Series and is the answer not in the manual?

ITech IT8200 Series Specifications

General IconGeneral
BrandITech
ModelIT8200 Series
CategoryMeasuring Instruments
LanguageEnglish