IEEE-488 Common Commands
Copyright © ITECH Electronic Co., Ltd. 112
Chapter21 IEEE-488 Common Commands
IEEE-488 Common commands usually control all instrument functions, such as
reset, status, and synchronization. All common commands consist of a
three-letter mnemonic, and are preceded by an asterisk, E.g:*RST *IDN? *SRE 8.
*CLS
This command clears the error queue and the bits of the following registers:
Standard Event Register
Questionable Event Register
Status byte register
Syntax:
*CLS
Arguments:
None
*ESE
This command sets the bits in the Event Status Enable Register (ESER). The ESER
is an eight-bit register that determines which bits in the Standard Event Status
Register (SESR) will set the Event Summary Bit (ESB) in the Status Byte Register
(SBR).
Syntax
*ESE <NR1>
Arguments
0255
Default value
0
Example
*ESE 128
Query syntax
*ESE?
Returns
<NR1>
Related syntax
*ESR? *PSC *STB?
Bit definition of standard event enable register:

Table of Contents

Other manuals for ITech IT8200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8200 Series and is the answer not in the manual?

ITech IT8200 Series Specifications

General IconGeneral
BrandITech
ModelIT8200 Series
CategoryMeasuring Instruments
LanguageEnglish