SENSE Subsystem
Copyright © ITECH Electronic Co., Ltd. 54
SENSe:FILTer:LEVel <CPD>
This command is used to set and query the filter level of the meter.
Syntax:
SENSe:FILTer:LEVel <CPD>
Arguments
SLOW|MEDIUM|FAST
Query syntax::
SENSe:FILTer:LEVel?
Returns::
SLOW|MEDIUM|FAST
Example
SENS:FILT:LEV FAST
SENSe:EXTernal:SYNC[:STATe] <boolean>
This command is used to set and query the state of synchronization.
Syntax:
SENSe:EXTernal:SYNC[:STATe] <boolean>
Arguments
0|OFF|1|ON
Query syntax::
SENSe:EXTernal:SYNC[:STATe]?
Returns::
0|1
Example
SENS:EXT:SYNC 1
SENSe:EXTernal:SYNC:PHASe <NRf>
Set the angle difference between the instrument and the external signal frequency.
When the external frequency lock function is turned on, the output phase of the
instrument maintains a fixed angle difference with the external frequency.
Syntax:
SENSe:EXTernal:SYNC:PHASe <NRf>