CONFigurate IO Subsystem
Copyright © ITECH Electronic Co., Ltd. 19
[CONFigurable:]IO:TYPE <NR1>,<CPD>
This command is used to set and query the IO type.
Note:
1: LIVing|LATCh
2: PSCLear
3: PSSTate
4: SYIN|SYOut
5: OFFState
6: TIN1|TOUT1
7: TIN2|TOUT2
Only specific IO can be configured for the above special functions. IORD|IOWR
all IO can be configured.
Syntax
[CONFigurable:]IO:TYPE <NR1>,<CPD>
Argument
NR1: [1,7]
CPD:
LIVing|LATCh|PSCLear|PSSTate|SYIN|SYOut|OFFState|TIN1|TIN2|TOUT1|TOU
T2|IORD|IOWR
Query syntax
[CONFigurable:]IO:TYPE? <NR1>
Returns
CPD
Example
IO:TYPE 1,LIV
[CONFigurable:]IO:TOUT:SOURce <CPD1>,<CPD2>
This command is used to set and query the trig source of IO-6 and IO-7 and
whether the function is on or off.
Note: this command should be used in conjunction with IO:SELect, that is, select
IO-6 or IO-7, and then set the trigger source.
Syntax
IO:TOUT:SOURce <CPD1>,<CPD2>
Argument
CPD1:AC|DC|LIST
CPD2:0|OFF|1|ON