Essential Commands
Copyright © ITECH Electronic Co., Ltd. 26
STATus:QUEStionable:NTRansition
The command set or read queries the value of the negative change enable
register, when the query register bit changes from 1 to 0, and the corresponding
bit of positive change enable register is 1, then the corresponding bit of the
quest event register is 1.
Command Syntax
STATus:QUEStionable:NTRansition <NR1>
Parameters
0 to 65535
Default value
0
Examples
STAT:QUES:NTRansition 32 STAT:QUES:
NTRansition 1
Query Syntax
STATus:QUEStionable:NTRansition?
Returned Parameters
<NR1> (register value)
Related Commands
STAT:QUES?
STATus:QUEStionable:CONDition?
This command can read the parameter from quest condition register. It is a read
only register, keep the real-time (not locked) query status of the load.
Query Syntax
STATus:QUEStionable:CONDition?
Parameters
None
Examples
STAT:QUES:COND?
Returned Parameters
<NR1> (register value)
Related Commands
STAT:OPER:COND?
STATus:OPERation?
This command query the query operation event register values. The event
register is read-only register, which holds (latches) all value passed by the NTR
and, or PTR filter. Read channel operation event register will clear it.
Query Syntax
STATus:OPERation[:EVENt]?
Parameters
None
Examples
STAT:OPER:EVEN?
Returned Parameters
<NR1> (register value)
Related Commands
*CLS
STATus:OPERation:ENABle
The command and its query set and read the parameters of operations enable
register. This register is a mask for enabling specific bits from the Operation
Event register to set the operation summary bit (OPER) of the Status Byte
register. The operation summary bit is the logical OR of all enabled Operation
Event register bits.
Command Syntax
STATus:OPERation:ENABle <NR1>
Parameters
0 to 255
Default value
0
Examples
STAT:OPER:ENAB 32 STAT:OPER:ENAB 1

Table of Contents

Other manuals for ITech IT8800 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8800 Series and is the answer not in the manual?

ITech IT8800 Series Specifications

General IconGeneral
BrandITech
ModelIT8800 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals