Calibration Commands
Copyright © ITECH Electronic Co., Ltd. 61
Chapter12 Calibration Commands
CALibration Subsystem
Calibration commands’ function:
Enable or disable calibration mode
Calibrate input function, current offset or gain, and save the newest
calibration constant in nonvolatile memory.
CALibrate:SECure[:STATe]
This command can enable or disable the calibration mode. You must enable
calibration mode so that you can receive other calibration commands. The first
parameter defines enable or disable status. The second parameter represents
the code. If the calibration mode is enabled while the current code is not 0,in
this circumstance, you need to configure the second parameter. If the code is
not set or incorrect, a error will be generated and calibration mode remains
disabled. Query command can only return the status of calibration mode but
code. Everytime, when calibration status changed from enabled status to
disabled status, any new calibration constants will be lost after power-down
unless you have utilized the save command CALibrate:SAVE.
Command Syntax CALibrate:SECure[:STATe] <bool> [,<SRD>]
Parameters 0 | 1 | OFF | ON [,<password>]
*RST Value ON
Examples CAL:SEC 0, N3301A CAL:SEC ON
Query Syntax CALibrate:SECure[:STATe]?
Returned Parameters <NR1>
Related Commands CAL:SAVE CAL:INIT
CALibrate:INITial
This command only be used in calibration mode. It re-saves the factory
calibration constants in nonvolatile memory.
Command Syntax CALibrate:INITial
Parameters None
Examples CAL:INIT
Related Commands CAL:STAT CAL:INIT
CALibrate:SAVe
This command only be used in calibration mode. Save the new calibration
constants in nonvolatile memory(after finishing the voltage or current calibration
process).
Command Syntax CALibrate:SAVE
Parameters None
Examples CAL:SAVE
Related Commands CAL:STAT CAL:INIT
CALibrate:CURRent:POINt
This command is used to set the calibration point in CC mode, only effective in

Table of Contents

Other manuals for ITech IT8800 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8800 Series and is the answer not in the manual?

ITech IT8800 Series Specifications

General IconGeneral
BrandITech
ModelIT8800 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals