Source Subsystem
Copyright © ITECH Electronic Co., Ltd. 51
[SOURce:]VOLTage:HIGH
[SOURce:]VOLTage:LOW
This command sets the high and low level of the current when in constant
voltage mode.
Command Syntax [SOURce:]VOLTage:HIGH <NRf+>
[SOURce:]VOLTage:LOW <NRf+>
Parameters MINimum through MAX | MINimum | MAXimum |
DEFault
Unit A (amps)
*RST Value MAXimum
Examples VOLT:HIGH 5
Query Syntax [SOURce:]VOLTage:HIGH?
[MINimum|MAXimum|DEFault ]
[SOURce:]VOLTage:LOW?
[MINimum|MAXimum|DEFault ]
Returned Parameters <NR3>
[SOURce:]RESistance
This command sets the resistance of the load when operating in constant
resistance mode.
Command Syntax [SOURce:]RESistance[:LEVel][:IMMediate] <NRf+>
Parameters MINimum through MAX | MINimum | MAXimum |
DEFault
Unit
*RST Value MAXimum
Examples RES 5 RES:LEV 3.5
Query Syntax [SOURce:]RESistance[:LEVel][:IMMediate]?
[ MINimum | MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands RES:RANG
[SOURce:]RESistance:RANGe
This command sets the resistance range of the electronic load module.
Command Syntax [SOURce:]RESistance:RANGe <NRf+>
Parameters MINimum through MAX | MINimum | MAXimum |
DEFault
Unit
*RST Value MAXimum (high range)
Examples RES:RANG 15 SOUR:RES:RANGE MIN
Query Syntax [SOURce:]RESistance:RANGe? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
[SOURce:]RESistance:TRANsient:MODE
This command selects the operating mode of the transient generator as follows
in constant resistance mode.