Calibration Commands
Copyright © ITECH Electronic Co., Ltd. 62
calibration mode.P1,P2 are used in low current range. P3, P4 are used in high
current range.
Command Syntax CALibrate:CURRent:POINt <point>
Parameters P1 | P2 | P3 | P4
Examples CAL:CURR:POIN P2
Related Commands CAL:STAT CAL:SAV
CALibrate:CURRent[:LEVel]
This command is only used in calibration mode. Input the calibration current
value according to the external measuring meter. But you must select a
calibration grade as a reference to the value you just input. These constants do
not exist in the nonvolatile memory before you send the save command
CALibrate:SAVE.
Command Syntax CALibrate:CURRent[:LEVel] <NRf>
Parameters <external reading>
Unit A (amps)
Examples CAL:CURR 3.2223
Related Commands CAL:STAT CAL:SAV
CALibrate:CURRent:METEr:POINt
This command is used to set the calibration point in CC mode, only effective in
calibration mode.P1,P2 are used in low current range. P3, P4 are used in high
current range. This can only calibrate the ammeter that selected through
command CONF:CURR.
Command Syntax CALibrate:CURRent:METEr:POINt <point>
Parameters P1 | P2 | P3 | P4
Examples CAL:CURR:METEr:POIN P2
Related Commands CAL:STAT CAL:SAV
CALibrate:CURRent:METEr[:LEVel]
This command is only effective in calibration mode. Input the calibration current
value according to reading of external measuring meter. Before you input this
value, please select a calibration grade firstly (using the command
CAL:CURR:METE:POIN).These constant do not exist in the nonvolatile
memory unless you send the command CALibrate:SAVE.
Command syntax CALibrate:CURRent:METEr[:LEVel] <NRf>
Parameters <external reading>
Unit A (amps)
Examples CAL:CURR 3.2223
Related Commands CAL:STAT CAL:SAV
CALibrate:VOLTage:POINt
This command is only effective in calibration mode, is used to set the calibration
point in CV mode.P1, P2 are used in low voltage range,P3,P4 are used in high
voltage range. Only be used to calibrate voltage source and voltmeter.
Command Syntax CALibrate:VOLTage:POINt <point>
Parameters P1 | P2 | P3 | P4
Examples CAL:VOLT:POIN P2
Related Commands CAL:STAT CAL:SAV