Source Subsystem
Copyright © ITECH Electronic Co., Ltd. 55
*RST Value ALEVEL MAXimum , BLEVel MINnum
Examples POW:TRAN:ALEV 5 POW:TRAN:BLEV 0.5
Query Syntax [SOURce:] POWer:TRANsient:ALEVel? [ MINimum |
MAXimum | DEFault ]
[SOURce:] POWer:TRANsient:BLEVel? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands POW
[SOURce:]POWer:TRANsient:AWIDth
[SOURce:]POWer:TRANsient:BWIDth
This command specifies the transient pulse width of the input power.
Command Syntax [SOURce:] POWer:TRANsient:AWIDth <NRf+>
[SOURce:] POWer:TRANsient:BWIDth <NRf+>
Parameters 50 to 3600 us
Unit S (second)
*RST Value 1000uS
Examples POW:TRAN:AWID 0.001 POW:TRAN:BLEV 0.02
Query Syntax [SOURce:] POWer:TRANsient:AWIDth? [ MINimum |
MAXimum | DEFault ]
[SOURce:] POWer:TRANsient:BWIDth? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands POW
[SOURce:]POWer:HIGH
[SOURce:]POWer:LOW
This command sets the voltage high and low level in constant power mode.
Command Syntax [SOURce:]POWer:HIGH <NRf+>
[SOURce:]POWer:LOW <NRf+>
Parameters MINimum through MAX | MINimum | MAXimum |
DEFault
Unit V (volts)
*RST Value MAXimum
Examples POW:HIGH 5
Query Syntax [SOURce:]POWer:HIGH?
[MINimum|MAXimum|DEFault ]
[SOURce:]POWer:LOW?
[MINimum|MAXimum|DEFault ]
Returned Parameters <NR3>
[SOURce:]POWer:PROTection
This command sets the soft power protection level. If the input power exceeds
the soft power protection level for the time specified by POW:PROT:DEL, the
input is turned off.