Source Subsystem
Copyright © ITECH Electronic Co., Ltd. 50
Command Syntax [SOURce:] VOLTage:TRANsient:MODE
<mode>
Parameters CONTinuous | PULSe | TOGGle
*RST Value CONTinuous
Examples VOLT:TRAN:MODE TOGG
Query Syntax [SOURce:] VOLTage:TRANsient:MODE?
Returned Parameters <CRD>
Related Commands VOLT:TRAN:ALEV TRAN
[SOURce:]VOLTage:TRANsient:ALEVel
[SOURce:]VOLTage:TRANsient:BLEVel
This command specifies the transient level of the input voltage. The transient
function switches between the level a and level b.
Command Syntax [SOURce:] VOLTage:TRANsient:ALEVel <NRf+>
[SOURce:] VOLTage:TRANsient:BLEVel <NRf+>
Parameters MIN through MAX | MINimum | MAXimum | DEFault
Unit V (volts)
*RST Value ALEVEL MAXimum , BLEVel MINnum
Examples VOLT:TRAN:ALEV 5 VOLT:TRAN:BLEV 0.5
Query Syntax [SOURce:] VOLTage:TRANsient:ALEVel?
[ MINimum | MAXimum | DEFault ]
[SOURce:] VOLTage:TRANsient:BLEVel?
[ MINimum | MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands VOLT
[SOURce:]VOLTage:TRANsient:AWIDth
[SOURce:]VOLTage:TRANsient:BWIDth
This command specifies the transient pulse width of the input voltage.
Command Syntax [SOURce:] VOLTage:TRANsient:AWIDth <NRf+>
[SOURce:] VOLTage:TRANsient:BWIDth <NRf+>
Parameters 50 to 3600 us
Unit S (second)
*RST Value 1000uS
Examples VOLT:TRAN:AWID 0.001 VOLT:TRAN:BLEV 0.02
Query Syntax [SOURce:] VOLTage:TRANsient:AWIDth?
[ MINimum | MAXimum | DEFault ]
[SOURce:] VOLTage:TRANsient:BWIDth?
[ MINimum | MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands VOLT