Source Subsystem
Copyright © ITECH Electronic Co., Ltd. 49
Examples VOLT:RANG 15
Query Syntax [SOURce:]VOLTage:RANGe? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands VOLT
[SOURce:]VOLTage:RANGe:AUTO[:STATe]
This command sets the voltage range of the electronic load module.
Command Syntax [SOURce:]VOLTage:RANGe:AUTO[:STATe] <bool>
Parameters 0 | 1 | ON | OFF
*RST Value 1
Examples VOLT:RANG:AUTO 1
Query Syntax [SOURce:]VOLTage:RANGe:AUTO[:STATe]?
Returned Parameters <NR1>
[SOURce:]VOLTage:ON
This command sets the voltage of sink current on.
Command Syntax [SOURce:]VOLTage[:LEVel]:ON <NRf+>
Parameters 0 through MAX | MINimum | MAXimum | DEFault
Unit V (volts)
*RST Value MINimum
Examples VOLT:ON 5
Query Syntax [SOURce:]VOLTage[:LEVel]:ON? [ MINimum |
MAXimum | DEFault ]
Returned Parameters <NR3>
Related Commands VOLT:LATCh
[SOURce:]VOLTage:LATCh
This command sets the action type of Von.
Command Syntax [SOURce:]VOLTage:LATCh[:STATe] <b>
Parameters 0 | 1 | ON | OFF
*RST Value ON
Examples VOLT:LATC 1
Query Syntax [SOURce:]VOLTage:LATCh[:STATe]? [ MINimum |
MAXimum | DEFault ]
Returned Parameters 0 | 1
Related Commands VOLT:ON
[SOURce:]VOLTage:TRANsient:MODE
This command selects the operating mode of the transient generator as follows
in constant voltage mode.
CONTinuous The transient generator puts out a continuous pulse stream
after receipt of a trigger signal.
PULSe The transient generator puts out a single pulse upon receipt of
a trigger.
TOGGle The transient generator toggles between two levels upon