Source Subsystem
Copyright © ITECH Electronic Co., Ltd. 46
[SOURce:]CURRent:SLEWrate:STATe
This command enables or disables the over-current slow rise mode.
Command Syntax [SOURce:]CURRent:SLEWrate[:STATe] <Bool>
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples CURR:SLEW:STAT 1
Query Syntax [SOURce:]CURRent:SLEWrate:STATe?
Returned Parameters <NR3>
Related Commands CURR:SLEW
[SOURce:]CURRent:PROTection:STATe
This command enables or disables the over-current protection feature.
Command Syntax [SOURce:]CURRent:PROTection:STATe <Bool>
Parameters 0 | 1 | OFF | ON
*RST Value OFF
Examples CURR:PROT:STAT 1
Query Syntax [SOURce:]CURRent:PROTection:STATe?
Returned Parameters <NR3>
Related Commands CURR:PROT
[SOURce:]CURRent:PROTection
This command sets the soft current protection level. If the input current exceeds
the soft current protection level for the time specified by CURR:PROT:DEL, the
input is turned off.
NOTE: Use CURR:PROT:DEL to prevent momentary current limit conditions
caused by programmed changes from tripping the overcurrent protection.
Command Syntax [SOURce:]CURRent:PROTection:LEVel <NRf+>
Parameters 0 through MAX
Unit A (amperes)
*RST Value MAXimum
Examples CURR:PROT 2
Query Syntax [SOURce:]CURRent:PROTection:LEVel?
Returned Parameters NR3
Related Commands CURR:PROT:DEL CURR:PROT:STAT
[SOURce:]CURRent:PROTection:DELay
This command specifies the time that the input current can exceed the
protection level before the input is turned off.
Command Syntax [SOURce:]CURRent:PROTection:DELay
<NRf+>
Parameters 0 to 60 seconds | MINimum | MAXimum | DEFault
Unit seconds
*RST Value 3
Examples CURR:PROT:DEL 5
Query Syntax [SOURce:]CURRent:PROTection:DELay?
[ MINimum | MAXimum | DEFault ]