IEEE-488 Reference
3-21
3.10 Programming syntax
The following information covers syntax for both common commands and SCPI commands.
For information not covered here, refer to the IEEE- 488.2 and SCPI standards.
Command words
Program messages are made up of one or more command words.
1. Commands and command parameters: Common commands and SCPI commands may
or may not use a parameter. Examples:
*SAV <NRf> Parameter (NRf) required.
*RST No parameter used.
:INITiate:CONTinuous <b> Parameter (<b>) required.
:SYSTem:PRESet No parameter used.
Note that there must be at least one space between the command word and the parameter.
Brackets [ ]: There are command words that are enclosed in brackets ([]). These brackets
are used to denote an optional command word that does not need to be included in the pro-
gram message. For example:
:INITiate[:IMMediate]
The brackets indicate that :IMMediate is implied (optional) and does not have to be used.
Thus, the above command can be sent in one of two ways:
:INITiate or :INITiate:IMMediate
Notice that the optional command is used without the brackets. Do not include the brack-
ets when using an optional command word.
Parameter types: Some of the more common parameter types are explained as follows:
<b> Boolean: Used to enable or disable an instrument operation. 0 or OFF dis-
ables the operation, and 1 or ON enables the operation. Example:
:CURRent:DC:RANGe:AUTO ON Enable auto-ranging.
<name> Name parameter: Select a parameter name from a listed group. Example:
<name> = NEVer
= NEXt
= ALWays
= PRETrigger
:TRACe:FEED:CONTrol PRETrigger
<NRf> Numeric representation format: This parameter is a number that can be ex-
pressed as an integer (e.g., 8), a real number (e.g., 23.6) or an exponent
(2.3E6). Example:
:STATus:MEASurement:ENABle Set bit B2 of enable register

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals