IEEE-488 Reference
3-134
:ZCORrect:ACQuire
:SYSTem:ZCORrect:ACQuire Acquire zero correct value
Description This command is used to acquire a zero correct value for zero correction. The value can only be
acquired while zero check is enabled. The internal offset will become the correction value.
Zero correction can then be performed with zero check disabled. The description for :ZCOR-
rect[:STATe] provides a command sequence using an acquired zero correction value.
3.22.13 A/D Controls
:TSControl <b>
:SYSTem:TSControl <b> Control external temperature readings
Parameters <b> = 0 or OFF Disable external temperature readings
1 or ON Enable external temperature readings
Query :TSControl? Query state of temperature readings
Description This command is used to control external temperature readings. When enabled (and the Model
6517-TP connected to the instrument), the Model 6517A will make external temperature read-
ings. Note that ETEMperature has to be a selected data element in order to include the temper-
ature reading in the data string. See paragraphs 3.17 (:FORMat:ELEMents) and 3.23
(:TRACe:ELEMents).
:HSControl <b>
:SYSTem:HSControl <b> Control humidity readings
Parameters <b> = 0 or OFF Disable humidity readings
1 or ON Enable humidity readings
Query :HSControl? Query state of humidity readings
Description This command is used to control humidity readings. When enabled (and the Model 6517-RH
connected to the instrument), the Model 6517A will make humidity readings. Note that HUMid-
ity has to be a selected data element in order to include the humidity reading in the data string.
See paragraphs 3.17 (:FORMat:ELEMents) and 3.23 (:TRACe:ELEMents).
:HLControl <b>
:SYSTem:HLControl <b> Control A/D hardware limit
Parameters <b> = 0 or OFF Disable A/D hardware limit
1 or ON Enable A/D hardware limit
Query :HLControl? Query state of A/D hardware limit
Description This command is used to control the A/D hardware limit. With the hardware limit ON, the “Out-
OfLimit” error message is enabled to indicate that an out of range noise spike has occurred.
Disabling the hardware limit disables the error message. With the hardware limit OFF, noise
spikes may cause the reading to be erroneous.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals