IEEE-488 Reference
3-72
3.15 :DISPlay subsystem
The display subsystem controls the display of the Model 6517A and is summarized in Table 3-6.
:TEXT commands
:DATA <a>
:DISPlay[:WINDow[1]]:TEXT:DATA <a> Define message; top display
:DISPlay:WINDow2:TEXT:DATA <a> Define message; bottom display
Parameters <a> = ASCII characters for message
Types: String 'aa...a' or "aa...a"
Indefinite Block #0aa...a
Definite Block #XYaa...a
where; Y = number of characters in message:
Up to 20 for top display
Up to 32 for bottom display
X = number of digits that make up Y (1 or 2)
Query :DATA? Query the defined text message
Description These commands define text messages for display. A message can be as long as 20 characters
for the top display, and up to 32 characters for the bottom display. A space is counted as a char-
acter. Excess message characters results in an error.
An indefinite block message must be the only command in the program message or the last com-
mand in the program message. If you include a command after an indefinite block message (on
the same line), it will be treated as part of the message and is displayed instead of executed.
:STATe <b>
:DISPlay[:WINDow[1]]:TEXT:STATe <b> Control message; top display
:DISPlay:WINDow2:TEXT:STATe <b> Control message; bottom display
Parameters <b> = 0 or OFF Disable text message for specified display
1 or ON Enable text message for specified display
Query :STATe? Query state of message mode for specified display
Description These commands enable and disable the text message modes. When enabled, a defined message
is displayed. When disabled, the message is removed from the display.
GPIB Operation — A user defined text message remains displayed only as long as the instru-
ment is in remote. Taking the instrument out of remote (by pressing the LOCAL key or sending
LOCAL 27), cancels the message and disables the text message mode.
RS-232 Operation — A user defined test message can only be cancelled by using this command
to disable the message or by cycling power.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals