IEEE-488 Reference
3-60
:TSEQuence
:VRESistivity
:STIME <NRf>
:STIME?
:MVOLtage <NRf>
:MVOLtage?
:MTIMe <NRf>
:MTIMe?
:DTIMe <NRf>
:DTIMe?
:ALTPolarity
:OFSVoltage <NRf>
:OFSVoltage?
:ALTVoltage <NRf>
:ALTVoltage?
:MTIMe <NRf>
:MTIMe?
:DISCard <NRf>
:DISCard?
:READings <NRf>
:READings?
:SIResistance
:SVOLtage <NRf>
:SVOLtage?
:STIME <NRf>
:STIME?
:MVOLtage <NRf>
:MVOLtage?
:MTIMe <NRf>
:MTIMe?
:SQSWeep
:HLEVel <NRf>
:HLEVel?
:HTIMe <NRf>
:HTIMe?
:LLEVel <NRf>
:LLEVel?
:LTIMe <NRf>
:LTIMe?
:COUNt <NRf>
:COUNt?
:STSWeep
:STARt <NRf>
:STARt?
:STOP <NRf>
:STOP?
:STEP <NRf>
:STEP?
:STIME <NRf>
:STIME?
Specify bias time; 0 to 99999.9 (sec).
Query bias time.
Specify measure voltage; -1000 to 1000.
Query measure voltage.
Specify measure time; 0 to 9999.9 (sec).
Query measure time.
Specify discharge time; 0 to 99999.9 (sec).
Query discharge time.
Alternating polarity resistance/resistivity test path.
Specify offset voltage; -1000 to 1000.
Query offset voltage.
Specify alternating voltage; -1000 to 1000.
Query alternating voltage.
Specify measurement time; 0.5 to 9999.9 (sec).
Query measurement time.
Specify number of initial readings to discard; 0 to 9999.
Query number of initial readings to discard.
Specify number of readings to store. See paragraph 3.25.19.
Query number of readings to store.
Surface Insulation Resistance test path:
Specify bias voltage; -1000 to 1000.
Query bias voltage.
Specify bias time; 0 to 99999.9 (sec).
Query bias time.
Specify measure voltage; -1000 to 1000.
Query measure voltage.
Specify measure time; 0 to 9999.9 (sec).
Query measure time.
Square wave sweep test path:
Specify high level voltage; -1000 to 1000.
Query high level voltage.
Specify high level time; 0 to 9999.9 (sec).
Query high level time.
Specify low level voltage; -1000 to 1000.
Query low level voltage.
Specify low level time; 0 to 9999.9 (sec).
Query low level time.
Specify number cycles.
Query cycle count.
Staircase sweep test path:
Specify start voltage; -1000 to 1000.
Query start voltage.
Specify stop votlage; -1000 to 1000.
Query stop voltage.
Specify step voltage; -1000 to 1000.
Query step voltage.
Specify step time; 0 to 9999.9 (sec).
Query step time.
1 sec
+500V
1 sec
2 sec
0V
10V
15 sec
3
1
+50V
1 sec
+100V
1 sec
+1V
1 sec
-1V
1 sec
+1V
+10V
+1V
1 sec
3.25.7
3.25.10
3.25.11
3.25.8
3.25.17
3.25.18
3.25.11
3.25.19
3.25.20
3.25.6
3.25.7
3.25.10
3.25.11
3.25.12
3.25.13
3.25.14
3.25.15
3.25.16
3.25.2
3.25.3
3.25.4
3.25.7
Table 3-16 (Continued)
:TSEQuence command summary
Command Description
Default
parameter SCPI Ref.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals