IEEE-488 Reference
3-88
When the integration period is set using the :APERture command, the value for the :NPLCycle
command changes accordingly to reflect the new integration period. Conversely, if the integra-
tion period is set using the :NPLCycle command, the value for the :APERture command chang-
es accordingly.
On power-up, the instrument uses the NPLC value to determine the integration period. Thus, if
using a different power line frequency, NPLC will remain the same, but aperture may change.
Note: If line synchronization is enabled (see:LSYNc command in SYSTem subsystem), the in-
tegration period will not start until the beginning of the next power line cycle. For example, if a
reading is triggered at the positive peak of a power line cycle, the integration period will not start
until that power line cycle is completed. The integration period starts when the positive-going
sine wave crosses zero volts. See paragraph 2.12.8 for more details.
:AUTO <b>|ONCE
[:SENSe[1]]:VOLTage[:DC]:APERture:AUTO <b>|ONCE Control auto aperture; volts
[:SENSe[1]]:CURRent[:DC]:APERture:AUTO <b>|ONCE Control auto aperture; amps
[:SENSe[1]:RESistance:APERture:AUTO <b>|ONCE Control auto aperture; ohms
[:SENSe[1]]:CHARge:APERture:AUTO <b>|ONCE Control auto aperture; coulombs
Parameters <b> =1 or ONEnable auto aperture
0 or OFF Disable auto aperture
ONCE Enable and then disable auto aperture
Query :AUTO? Query state of auto aperture.
Description These commands are used to enable or disable auto aperture for the specified measurement
function. With auto aperture enabled, the instrument automatically optimizes the aperture value
for the present resolution setting. This is the same as selecting SET-BY-RSLN from the front
panel.
Table 2-17 provides the integration times for SET-BY-RSLN (AUTO). Note that the times are
provided as NPLC values. To convert to aperture times, use the formula in paragraph 3.19.4.
The ONCE parameter is analogous to a momentary toggle switch. When :AUTO ONCE is sent,
auto aperture turns on momentarily and then disables. While enabled, it automatically selects
the appropriate value as summarized in the tables. When auto aperture disables, the aperture re-
mains at the selected value.
:APERture:AUTO is coupled to the command that manually sets the aperture value (:APERture
<n>). When auto aperture is enabled, the parameter value for :APERture <n> changes to the au-
tomatically selected aperture value. Thus, when auto aperture is disabled, the aperture remains
at the automatically selected value. Also, a valid aperture value sent using the :APERture <n>
command disables auto aperture.
Auto aperture is also coupled to auto NPLC. Enabling or disabling auto aperture also enables or
disables auto NPLC. Also, a valid NPLC value sent using the :NPLC command disables auto
aperture.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals