IEEE-488 Reference
3-108
:LIMit:STATe <b>
:SOURce:VOLTage:LIMit:STATe <b> Control voltage limit
Parameters <b> = 0 or OFF Disable voltage limit
1 or ON Enable voltage limit
Query :STATe? Query state of voltage limit
Description This command is used to enable or disable voltage limit. When enabled, the programmed volt-
age limit is in effect. When disabled, voltage limit is determined by the selected range.
Attempts to enable voltage limit when the present V-Source level (amplitude) exceeds the limit
value will generate a settings conflict error.
:MCONnect <b>
:SOURce:VOLTage:MCONnect <b> Control LO connection
Parameters <b> = ON or 1 Enable LO connection
OFF or 0 Disable LO connection
Query :MCONnect? Query LO connection
Description This command is used to control the internal V-Source LO to Ammeter LO connection. When
enabled, V-Source LO is connected to Ammeter LO. This connection simplifies the connection
requirements for ohms measurements and source voltage measure current measurements. When
disabled, this connection is broken.
:CURRent Commands
:RLIMit:STATe <b>
:SOURce:CURRent:RLIMit:STATe <b> Control resistance current limit
Parameters <b> = 0 or OFF Disable resistance current limit
1 or ON Enable resistance current limit
Query :STATe? Query state of resistance current limit
Description This command is used to enable or disable resistance current limit for the V-Source. When en-
abled, a 1MΩ resistor is placed in series with the HI V-SOURCE OUT lead to extend the current
limit capabilities of the V-Source. When disabled, normal current limit occurs at 1mA.
:LIMit[:STATe]?
:SOURce:CURRent:LIMit[:STATe]? Query current compliance.
This command is used to determine if the V-Source has reached its current compliance limit of
1mA. The response message is explained as follows:
0 = Current compliance not reached
1 = Current compliance reached