IEEE-488 Reference
3-136
Parameters <name> = CONTinuous Continuous trigger mode
ONEShot One-shot trigger mode
Description This command is used to specify the basic trigger mode. With CONTinuous triggering, the Model
6517A will continuous trigger readings after the selected source event occurs (see :SOURce).
With ONEShot selected, a single reading will be triggered every time the selected source event
occurs.
:SOURce <name>
:SYSTem:MACRo:TRIGger:SOURce <name> Specify control source
Parameters <name> = IMMediate Immediate triggering
MANual Manual trigger event
BUS Bus trigger event
EXTernal External trigger event
TIMer Timer trigger event
Description This command is used to specify which control source will be used as the trigger event for instru-
ment readings. If in the continuous trigger mode (see :MODE), the instrument will continuously
make readings when the specified event occurs. If in the one-shot mode, each event will cause a
single-reading to occur.
With IMMediate selected, the trigger event will be satisfied immediately. In IMMediate, readings
will be made continuously.
With MANual selected, the trigger event occurs when the front panel TRIG key is pressed. Note
that the instrument must be in local in order for the TRIG key to be active.
With BUS selected, the trigger event occurs when a bus trigger (*TRG or GET) is sent. With EX-
Ternal selected, the trigger event occurs when an input trigger is received via the rear panel EXT
TRIG IN connector.
While in the one-shot mode, with TIMer selected, trigger events occur at a programmed timer in-
terval. Each time the programmed interval elapses, a reading is triggered. Use the :TIMer <n>
command to set the timer interval.
:TIMer <n>
:SYSTem:MACRo:TRIGger:TIMer <n> Set the timer interval
Parameters <n> = 0.001 to 99999.999 Specify timer interval in seconds
DEFault 0.1 seconds
MINimum 0.001 seconds
MAXimum 99999.999 seconds
This command is used to specify the time interval for the timer. The :SOURce TIMer command
is used to select the timer as the control source.

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals