EasyManuals Logo
Home>Keithley>Measuring Instruments>6517A

Keithley 6517A User Manual

Keithley 6517A
343 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #314 background imageLoading...
Page #314 background image
IEEE-488 BUS IMPLEMENTATION
MULTILINE COMMANDS: DCL, LLO, SDC, GET, GTL, UNT, UNL, SPE, SPD.
IMPLEMENTATION: SCPI (IEEE-488.2, SCPI-1993); DDC (IEEE-488.1).
UNILINE COMMANDS: IFC, REN, EOI, SRQ, ATN.
INTERFACE FUNCTIONS: SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E1.
PROGRAMMABLE PARAMETERS: Function, Range, Zero Check, Zero Suppress, EOI (DDC mode only), Trigger, Terminator (DDC mode only), 100-Reading Storage
(DDC mode), 15706 Max. Reading Storage (SCPI mode), Calibration (SCPI mode only), V-Source Output, Display Format, SRQ, Status (including V-Source I-Limit),
Output Format, Guard.
ADDRESS MODES: TALK ONLY and ADDRESSABLE.
TRIGGER TO READING DONE: 150ms typical, with external trigger.
RS-232 IMPLEMENTATION: Supports: SCPI 1991.0. Baud Rates: 300, 600, 1200, 2400, 4800, 9600, 19.2k.
PROTOCOLS: Xon/Xoff, 7- or 8-bit ASCII, parity-odd/even/none.
CONNECTOR: DB-9 TXD/RXD/GND.
GENERAL
DISPLAY: 6
1
2
-digit vacuum fluorescent multiline.
OVERRANGE INDICATION: Display reads “OVERFLOW.
RANGING: Automatic or manual.
CONVERSION TIME: Selectable 0.01 PLC to 10 PLC.
PROGRAMS: Provide front panel access to IEEE address, choice of engineering units or scientific notation, and digital calibration.
MAXIMUM INPUT: 250V peak, DC to 60Hz sine wave; 10s per minute maximum on mA ranges.
MAXIMUM COMMON MODE VOLTAGE (DC to 60Hz sine wave): Electrometer, 500V peak; V Source, 750V peak.
ISOLATION (Meter COMMON to chassis): Typically 10
10
in parallel with 500pF.
INPUT CONNECTOR: Three lug triaxial on rear panel.
2V ANALOG OUTPUT: 2V for full range input. Non-inverting in Volts and Ohms modes, inverting when measuring Amps or Coulombs. Output impedance 10k.
PREAMP OUTPUT: Provides a guard output for Volts measurements. Can be used as an inverting output or with external feedback in Amps and Coulombs modes.
EXTERNAL TRIGGER: TTL compatible External Trigger and Electrometer Complete.
GUARD: Switchable voltage guard available.
DIGITAL I/O AND TRIGGER LINE: Available, see manual for usage.
EMI/RFI: Meets VDE-0871 and FCC Class B limits.
EMC: Conforms to European Union Directive 89/336/EEC.
SAFETY: Conforms to European Union Directive 73/23/EEC (meets EN61010-1/IEC 1010).
TEST SEQUENCES: Device-Characterization (Diode, Capacitor, Cable, Resistor), Resistivity, Surface-Insulation-Resistance, Sweep.
READING STORAGE: 100 readings (DDC mode), 15706 max. readings (SCPI mode).
READING RATE:
To internal buffer 125 readings/second
1
To IEEE-488 bus 115 readings/second
1,3
To front panel 17 readings/second
2
Bus transfer 2500 readings/second
3
1
0.01 PLC, digital filters off, front panel off, temperature + RH off.
2
1.00 PLC, digital filters off, temperature + RH off.
3
Binary transfer mode.
DIGITAL FILTER: Median and averaging.
ENVIRONMENT: Operating: 0°–50°C; relative humidity 70% non-condensing, up to 35°C. Storage: –25° to +65°C.
WARM-UP: 1 hour to rated accuracy (see manual for recommended procedure).
POWER: 105–125V or 210–250V (external switch selected), 90–110V (internal switch selected), 50–60Hz, 50VA.
PHYSICAL: Case Dimensions: 90mm high × 214mm wide × 369mm deep (3
1
2
in. × 8
1
2
in. × 14
1
2
in.).
Working Dimensions: From front of case to rear including power cord and IEEE-488 connector: 15.5 inches.
Net Weight: <4.6 kg (<10.1 lbs).
Shipping Weight: <9.5 kg (<21 lbs).
Specifications subject to change without notice.
A-4 Specifications

Table of Contents

Other manuals for Keithley 6517A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517A and is the answer not in the manual?

Keithley 6517A Specifications

General IconGeneral
BrandKeithley
Model6517A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals