6-62 Maintenance
6.10.12 Bandwidth Problem
T
HIS
PROCEDURE
ASSUMES
THAT
THE
SYSTEM
IS
CAPABLE
OF
PASSING
A
SIGNAL
ON
EACH
CHANNEL
AND
THAT
IT
IS
NOT
DISJOINTED
OR
EXCESSIVELY
NOISY
I
S
THE
BANDWIDTH
ACCEPTABLE
?
Y
ES
N
O
P
ERFORM
F
RONT
E
ND
T
RIM
P
ROCEDURE
SEC
6.9.3.2.6
I
NPUT
A
100 K
HZ
SIGNAL
INTO
100
M
V/
DIV
. S
WEEP
THE
FREQUENCY
UP
TO
200MH
Z
, 350MH
Z
,
OR
500MH
Z
(
DEPENDING
ON
MODEL
).
I
S
THE
BANDWIDTH
ACCEPTABLE
?
Done
R
EPLACE
FE/ACQ
C
ARD
N
O
Yes