5-12 Performance Verification Rev. D Sept 2007
5.5.2 External Trigger Leakage Current
a. DC 50Ω
Recall Leakage - 50 ohm ext x1.lss or configure the DSO:
Panel Setups : Recall FROM DEFAULT SETUP
Channels Trace OFF Channel 1, Channel 2, Channel 3 & Channel 4
Select Setup trigger
Set Trigger on : EXT
Impedance : DC 50Ω
Connect the DMM to External.
Measure the voltage and enter it in Table 3. Compare it to the limits.
Recall Leakage - 50 ohm ext x10.lss or Set Trigger Source to Ext/10.
Repeat the test.
Record the measurements in Table 3, and compare the test results to the
limits in the test record.