3.10 Environmental and Safety
Temperature (Operating) +5 °C to +40 °C
Temperature (Non-Operating) –20 °C to +60 °C
Humidity (Operating) 5% to 80% relative humidity (non-condensing) at <= 30 °C.Upper limit
derates to 55% relative humidity (non-condensing) at +40 °C.
Humidity (Non-Operating) 5% to 95% relative humidity (non-condensing) as tested per MIL–PRF–
28800F.
Altitude (Operating) up to 3048 m (10,000 ft) at up to 25 °C
Altitude (Non-Operating) up to 12,190 m (40,000 ft)
Vibration (Operating) Random vibration, 0.31 grms 5 Hz to 500 Hz, 15 minutes in each of three
orthogonal axes
Vibration (Non-Operating) Random vibration, 2.4 grms 5 Hz to 500 Hz, 15 minutes in each of three
orthogonal axes
Functional Shock 20 g peak, half sine, 11 ms pulse, 3 shocks (positive and negative) in
each of three orthogonal axes, 18 shocks total
Certification CE Approved, UL (Std. UL 3111-1) and cUL (Std. CSA C22.2 No. 1010-
1) listed. EMC Directive 89/336/EEC;
EN61326-1:1997+A1:1998+A2:2001. Low Voltage Directive 73/23/EEC;
EN 61010-1:2001 Product Safety (Installation Category II, Pollution
Degree 2, Protection Class 1)
3.11 Physical Dimensions
Dimensions (HxWxD) 260 mm x 340 mm x 152mm (10.25" x 13.4" x 6"). Excluding accessories
and projections.
Net Weight 6.8 kg (15 lbs). Excluding accessories.
3.12 General
Power (AC) 100–120 Vrms at 50/60/400 Hz; 200–240 Vrms at 50/60 Hz;Max.
Power Consumption: 170 VA
3.13 Warranty & Calibration
Warranty Three year warranty.
Calibration Calibration recommended yearly.
3-4 Specifications