Repeat the test for all input channels.
Recall Leakage - 1 Mohm x10.lss or set Input gain to 200 mV/div. on all 4
Channels
Repeat the test for all input channels. Record the measurements in Table 3,
and compare the results to the limits in the test record.
Recall Leakage - 1 Mohm x1.lss or set Input gain to 2 V/div. on all 4 Channels
Repeat the test for all input channels. Record the measurements in Table 3,
and compare the results to the limits in the test record.
Rev. D Sept 2007 Performance Verification 5-11