A7™ Operating Instructions 046-004667-00 5 - 3
Operations Patient Setup
5.3 Patient Setup
5.3.1 End Case / Standby Mode
The End Case button is located in the Manual tab (see FIGURE 5-2). The End Case button is only
available when the Auto/Manual ventilation switch is set to Manual.
FIGURE 5-2 End Case Button
After selecting the End Case button, a warning box with a Restore default settings checkbox
will be displayed. Selecting the Restore default settings checkbox reloads the user defaults,
clears the patient demographics, the history, the spirometry reference loops, and places the
system into the Standby mode (see FIGURE 5-4).
If the Restore default settings checkbox is not selected, all the settings are retained.
FIGURE 5-3 End Case Checkbox
In the Standby mode, all system functions are not working. It is the default system startup mode
and is used after ending the case.