EasyManuals Logo
Home>Mindray>Medical Equipment>A7

Mindray A7 Operating Instructions

Mindray A7
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #215 background imageLoading...
Page #215 background image
A7™ Operating Instructions 046-004667-00 7 - 17
Maintenance Cleaning and Disinfection
7.9.7 PAW Gauge
1. The PAW gauge is a component that is plugged into position for use. It is not necessary to
remove this component to clean it. However, if removal is desired, simply hold it and lift it
straight up from the breathing system block (see FIGURE 7-14).
FIGURE 7-14 PAW Gauge Removal
CAUTION: Do not autoclave the following components: Paw gauge, flow sensor,
and bellows. These components cannot withstand immersion or the
heat and pressure of autoclaving.
2. Clean the PAW gauge with a soft, lint-free cloth and a recommended cleaning agent (see
TABLE 7-2 on page 7-8 and TABLE 7-3 on page 7-9). Allow it to dry thoroughly.
3. Re-insert the PAW gauge if it was removed. Prior to use after cleaning or disinfecting, power
up the system and follow the on-screen prompts to perform the leak test and the
compliance test (see section 4.5 (page 4-8) "Leak and Compliance Tests").

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals