Alarm and Prompt Messages Alarms and Messages
6 - 28 046-004667-00 A7™ Operating Instructions
6.6.3 Prompt Messages
6.6.3.1 Prompt Messages Displayed in Alarm Area
MESSAGE REMARK
Pressure, Volume and
Apnea Alarms are OFF
This message displays when the Alarms button in the Manual mode tab is
set to Off.
CO
2
and CO
2
Apnea
Alarms are OFF
This message appears when the CO
2
Alarms button in the Manual mode tab
is set to Off.
Load Configuration
Failure
This message displays when the download or latest configuration update
failed.
DEMO Mode - Not for
Clinical Use
This message displays when the system is set to demo mode from the
Service tab.
Service Mode - Not for
Clinical Use
This message displays when the machine is worked in Service mode.
Apnea Ventilation This message displays when apnea ventilation is triggered in PS mode.
Auto-zero in process This message displays when auto-zeroing of the pressure sensors is in
process.
New functions
activated, please
restart!
This message displays when activation successfully completed.
Restart to Activate
New Flowmeter
Standard
This message displays when flowmeter standard is changed.
Calibrate internal AG This message displays when the external AG module hasn't calibrated
successfully for 365 days.
Calibrate external AG This message displays when the external AG module hasn't calibrated
successfully for 365 days.
Could not locate time
server
This message displays when the Interval of SNTP Protocol is not Off and the
time server is unavailable for 5 intervals.
External AG Loaded
Successfully.
External AG loaded successfully.
External AG Unloaded
Successfully.
External AG unloaded successfully.
External AG Startup External AG module is starting up.
External AG Warmup External AG module is warming up.
External AG Zeroing The external AG module is being zeroed.
Leak Test Not
Performed
This message displays when either the automatic leak test or manual leak
test was skipped from startup, or when the last time that the leak test was
performed was more than 24 hours ago.
Ventilation and Fresh
Gas Flow Paused
This message displays when the Flow Pause is active.
All Physiological
Alarms are OFF
This message displays when the Flow Pause is active.
TABLE 6-10 Prompt Messages Display in Alarm Area

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals