Physical Views Product Description
1 - 18 046-004667-00 A7™ Operating Instructions
1.2.7 Breathing System (Left View)
FIGURE 1-8 Breathing System (Left View, the Flexible Bag Arm (optional))
F7 Quick Release APL Valve
2
Rotary regulator for setting the breathing system pressure
limit during manual ventilation. Its scale shows approximate
pressure. Set to SP during Spontaneous breathing. When
necessary, lift the APL valve upward to release pressure
quickly.
F8 Auto/Manual ventilation
Switch
Enables switching between Automatic and Manual
ventilation modes
PART(S) DESCRIPTION
1
The bellows dome is a transparent cover with graduation marks from 300 to 1500 ml. These marks are for ref-
erence only. Tidal volume (Vt) should be read exclusively from the display of the user interface. Delivered Vt is
a combination of bellows displacement and fresh gas flow.
2
The APL valve and PAW gauge numerics are for reference only. Calibrated patient airway pressure is dis-
played on the user interface.
G1
G2
G3
G4
G5
G6

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals