A7™ Operating Instructions 046-004667-00 xix
Notes Introduction
CAUTION: The AGSS three ways connector provides an inlet to the AGSS when the
ACGO circuit is used. Keep the inlet port cover closed when the ACGO
circuit is not used.
Notes
NOTE: Figures in this manual are provided for reference purposes only.
Screens may differ based on the system configuration and selected
parameters.
NOTE: Put the equipment in a location where you can easily see the screen and
access the operating controls.
NOTE: Keep this manual close to the equipment so that it can be obtained
conveniently when needed.
NOTE: The software was developed in compliance withANSI/ AAMI ES60601-1.
The possibility of hazards arising from software errors is minimized.
NOTE: This manual describes all features and options. Your equipment may
not have all of them.
NOTE: The A7 is intended to be operated with its integral Breathing Pressure
monitoring in use.
NOTE: The A7 is intended to be operated with its integral Breathing Pressure
limitation devices in use.
NOTE: The A7 is intended to be operated with its integral Exhaled Volume
monitoring in use.
NOTE: The A7 is intended to be operated with its integral Breathing System
integrity Alarm System in use.
NOTE: The A7 is intended to be operated with its integral Continuing Pressure
Alarm in use.
NOTE: The A7 is intended to be operated with its integral O
2
monitoring in
use.
NOTE: The A7 is intended to be operated with an external CO2 monitor
complying with ISO 80601-2-55. Connection to the CO2 monitor should
be via a sample line from the patient circuit.
NOTE: An Anesthesia Vapor Delivery Device is to be used with an Anesthetic
Agent Monitor complying with ISO 80601-2-55. Connection to the
Agent monitor should be via a sample line from the Patient Circuit.
NOTE: Continuously monitor the anesthetic agent concentration when using
the Anesthesia System to ensure accurate output of the anesthetic
agent.
NOTE: Check the liquid level of the anesthetic agent before and during all
operations. When the liquid level is below the warning line, more
anesthetic agent needs to be added. Refer to the vaporizer Instructions
For Use for filling the vaporizer and other information.
NOTE: The A7 System is designed to be equipped with an anesthetic vapor
delivery device that complies with ISO 80601-2-13.