A7™ Operating Instructions 046-004667-00 9 - 3
Product Specifications Safety Designations
• The pressure measurement device in compliance with ISO 80601-2-13;
• The pressure restriction device in compliance with ISO 80601-2-13;
• The expiratory volume monitor in compliance with ISO 80601-2-13;
• The breathing system with alarm system in compliance with ISO 80601-2-13;
• The anaesthetic vapour delivery system in compliance with ISO 80601-2-13;
• The anaesthetic gas scavenging system in compliance with ISO 80601-2-13;
• The anesthetic gas delivery device in compliance with ISO 80601-2-13;
• The anesthetic ventilator in compliance with ISO 80601-2-13;
•The O
2
monitor in compliance with ISO 80601-2-55;
•The CO
2
monitor in compliance with ISO 80601-2-55;
• The AG monitor in compliance with ISO 80601-2-55.
The anesthesia workstation is integrated with the pressure measurement device, pressure restriction
device, expiratory volume monitor, anaesthetic breathing system with alarm system, anaesthetic gas
delivery system, anaesthetic vapour delivery system, anaesthetic ventilator, AG monitor in
compliance with the afore mentioned standards, where:
• The pressure restriction device, expiratory volume monitor and breathing system with alarm
system also comply with ISO 80601-2-13.
• AG monitor in compliance with ISO 80601-2-55.
9.2 Safety Designations
Type of Protection against
Electric Shock:
Class I equipment with internal electric power supply.
Where the integrity of the external protective earth
(ground) in the installation or its conductors is in doubt, the
equipment shall be operated from its internal electric power
supply (i.e., battery supply).
Degree of Protection against
Electric Shock:
BF, defibrillation-proof
Power Supply Connection: External electric power supply:
100 to 120 VAC, 50/60 Hz, 12 A
Internal battery supply:
Lithium-ion, 11.1 V, 4.5 Ah ( 2 batteries installed)
Mode of Operation: Continuous
Degree of Protection against
Hazards of Explosion:
Ordinary equipment, without protection against explosion;
not for use with flammable anesthetics.
Degree of Protection against
Harmful Ingress of Water:
Protection against vertically falling water drops - IPX1 (IEC
60529)
Electrical Connection between
Equipment and Patient:
Equipment designed for non-electrical connection to the
patient
Degree of Mobility: Mobile: including the base and casters of the anesthesia
system
Disinfection: Steam autoclavable or disinfectable
TABLE 9-2 Safety Designations