9-4 Structure and Assembly/Disassembly
9.3.1 Display (Monitor) Assembly
The disassembly tool: cross-headed screwdriver (M3, M4)
1. Unscrew two M4X12 screws, and remove the display cable cover;
2. Unscrew three M4 X 12 screws. Remove the power supply cable of the display and unplug the
signal cable.
3. Unscrew six M4X12 screws, and remove the display assembly upwards.
NOTE:
The metal part is inserted in the hole of the plastic part. The damage of the display’s
rear cover may occur if removed by force.