4-14 Product Principle
4.4.5 Network Interface
Connect the network to ultrasound device to accomplish DICOM, remote control, remote print, etc.
Network capability of DC-80A product:
âš« Supports wired network: ten trillion Ethernet, hundred trillion Ethernet and gigabit Ethernet.
âš« Supports wireless network: Wi-Fi (802.11b/g/n)
Protocol:
â—¼ Wired network: is realized based on PC module (MAC+PHY) and isolation transformer of IO
board. The network capability is decided by PC module.
â—¼ Isolation transformer and RJ45 network circuit have to meet the security requirements of
creepage distance.
â—¼ Wireless network: uses built-in wireless module. The network capability is decided by wireless
module and cables. The regular wireless module only owns Wi-Fi. Purchase the wireless
module integrated with Bluetooth function. Wireless module is mini PCIe half-size card. Wi-Fi
on the wireless module communicates with PC module via PCIe x1, and Bluetooth
communicates with PC module via USB x1.
4.4.6 PCIE Interface Distribution
The data and controls of ultrasound devices are based on PCIE interface.
PGIE interface distribution plans of DC-80A product are:
4.4.7 Other IO Extension
IO extension of DC-80A product:
âš« Battery management.
âš« System monitoring (power supply, temperature, fan).
âš« PHV debugging
âš« LPC interface on PC module extends to connect to varied devices via MF_FPGA