9-6 Structure and Assembly/Disassembly
9.3.3 Disassembling ECG Assembly
The disassembly tool: cross-headed screwdriver (M3, M4), diagonal cutting pliers.
1. Refer Chapter 9.3.2 from step 1 to step 3. Remove the left and right side panels and the front
cover.
2. Unplug the signal cable connecting ECG assembly to mother board and MIC cable, and release
the clasp.
3. Unscrew three M4 X 8 screws fixing the assembly of front output panel, and pull the assembly
of front output panel out.
NOTE:
1.
Because of the socket of the probe board lying on the left side, please use even
strength when pulling out it.
2.
There are two bolts on the probe board assembly. Snap the assembly into place after
inserting the bolts, and then fix the screws.
ECG
MIC