System Diagnosis and Support 11-13
11.3.7 Other Errors
11.4 Self Test
11.4.1 Self Test Introduction
The self test function, adopted by DC-80A series products, is used to test the connection of
hardware board and running status of the device. According to the access authority and tests, there
are three types of tests: production self test, maintenance self test and user self test. This chapter
describes the maintenance self test and user self test in detail.
11.4.2 Operation Procedure of Maintenance Self Test
All tasks running on Doppler should be completed before entering system self test,
otherwise the self test system fails to respond. It is recommended to start self test
software to perform the self test after the system is generated.
1. Switch account, type service password and service account to log in.
2. Select [Setup]-[Maintenance]-[Setup].
Click [Self Test] out of dialog box.
NOTE: if the current Doppler system is in Chinese, the self test performs Chinese system self test.
NOTE: if the current Doppler system is not in Chinese, the self test performs English system self
test.
Fail to open the file
"SystemConfiguration.ini", and
please check HDD data!
Re-install the system software.
ADT7462 initialization error
XXX ADT7462
initialization errorï¼›
Of which the names
of XXX ADT 7462
are as follows:
PC Board A
PC Board B
DC-DC Board
Re-install the system
software/check the drive/check the
digital board/check the front-end
power main board.