9-30 Structure and Assembly/Disassembly
9.3.16 Face Cover Assembly of the Speaker/Speaker
The disassembly tool: cross-headed screwdriver (M3, M4).
1. Unscrew two M4 X 12 SEMS screws from the face cover assembly of the speaker. Lift the
assembly towards the arrow’s direction. Same procedures to remove right assembly.
2. Unscrew four M3 X 8 SEMS screws from the speaker to remove it. Disconnect the cables
connecting the speaker to the main unit to remove the speaker assembly. Same procedures to
remove right assembly.
NOTE:
The connecting cables lie behind the rear cover of the touchscreen. Be careful of
the cables when removing the assembly.
NOTE:
To avoid incorrect installation, please take notice of the position of the speakers.