Product Principle 4-7
◼ Response to the software control. The software configures DSP_FPGA’s register to complete
the configuration on entire front-end via PCIE interface.
â—¼ TR control board. XCVR from DSP_FPGA controls each TR board respectively.
â—¼ Scanning generator. It is produced by DSP_FPGA.
â—¼ The receiving, pro-processing, uploading of the beam data. XCVR from DSP_FPGA receives
beam data on TR board; logic completes the process and the calculation and sends back to PC
module via PGIE. Logic algorithm of the pro-processing can be upgraded.
â—¼ Clock distribution includes front-end clock and XGVR reference clock. The clock source and
distribution circuit is available inside the board. Distribute DSP_FPGA and each TR board.
â—¼ RATE distribution. RATE, a special clock, routes around with front-end clock and distributes to
DSP_FPGA and each TR board.
◼ ATGC distribution. Distribute ATGC’s differential voltage to each TR board.
â—¼ Probe management. DSP_FPGA completes the probe management via SPI controlling CPLD
on control probe board. CPLD online upgrade.
â—¼ The pencil probe is in the place and has the access to the ID.ID signal refers to one wire signal.
The specialized buffer is needed.
â—¼ Supports the physiological signal module and provides the serial.
â—¼ Distributes JTAG chain facing forwards the front-end.
◼ Provides 4D’s control and driving as the carrier of 4D&TEE board.
â—¼ Test signal of output acoustic power.
4.2.5 4D&TEE Module
The motor inside 4D probe requires the driving signal. With the specialized requirements of the
voltage and the phase of the driving signal, the driving circuit comes into form.
Temperature signal test on TEE probe needs balance bridge current and ADC sampling.
TEE probe and 4D probe are both optional. The related circuit has an independent module. The
supportive circuits of the two probe bound together come into 4D&TEE module and are placed
above the engine board.
HALL digital signal from 4D probe shows the location of the probe. The signal does not arrive at
4D&TEE module, but goes into DSP_FPGA directly.

Table of Contents

Other manuals for Mindray DC-80A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray DC-80A and is the answer not in the manual?

Mindray DC-80A Specifications

General IconGeneral
BrandMindray
ModelDC-80A
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals