10.7 AUDIO PAUSED
10.7.1 Set AUDIO PAUSED
Push the key to pause audio alarm of currently active alarms for 120 seconds.
WARNING
Pay close attention to the patient and ventilator to ensure no alarm messages are
ignored during the period of AUDIO PAUSED. Possible patient or equipment
hazard may be produced if the alarm condition continues while no action is taken.
NOTE
Under AUDIO PAUSED status, all the alarm indicators work normally except
audible alarm tones.
Under AUDIO PAUSED status, if a new technical or physiological alarm occurs,
the AUDIO PAUSED status terminates automatically and audible alarm tones
start again.
When the 120 s countdown time is up, the AUDIO PAUSED status terminates and
audible alarm tones start again.
10.7.2 Terminate AUDIO PAUSED
Under AUDIO PAUSED status, pushing the key or triggering a new alarm will
terminate the AUDIO PAUSED status and restore audible alarm tones. The AUDIO PAUSED
icon and 120s countdown will disappear from the screen at the same time.
10.8 Recent Alarm
When there are currently active alarms, if the number is displayed before alarm messages, it
indicates there are multiple all active alarm messages. By selecting the alarm message field,
you can view active alarm messages, alarm occurrence time and alarm priority in the
accessed most recent alarm window. Up to 9 alarm messages are displayed.
The icon
is displayed when all active alarms are cleared and there are no currently active
alarms. By pushing the icon
, you can view the most recent inactive alarms in the accessed
window (up to 9 alarm messages are displayed). You can also clear the most recent inactive
alarms with the
button.