5-6
NOTE
The system total response time for oxygen concentration monitoring is 23s.
It takes approximately 3 minutes from powering on the ventilator to reaching the
oxygen concentration monitoring performance specified in section B.7 of this
manual.
5.11 Set Language
1. Select [Setup]→[Maintain]→[User]→enter the required password→[Setting].
2. Select [Language] and select the desired language.
3. Restart the ventilator to activate the selected language.
5.12 Set Unit
5.12.1 Set Weight Unit
1. Select [Setup]→[Maintain]→[User]→enter the required password→[Setting]→
[Unit].
2. Set [Weight Unit] and toggle between [kg] and [lb].
5.12.2 Set Paw Unit
1. Select [Setup]→[Maintain]→[User]→enter the required password→[Setting]→
[Unit].
2. Select [Paw Unit] among [cmH
2
O], [hPa], and [mbar].
5.12.3 Set CO
2
Unit
1. Select [Setup]→[Maintain]→[User]→enter the required password→[Setting]→
[Unit].
2. Select [CO
2
Unit] among [mmHg], [kPa], and [%].
5.13 Set O
2
Supply Type
1. Select [Setup]→[Maintain]→[User]→enter the required password→[Setting]→[Gas
Supply].
2. Select [O
2
Supply Type] and toggle between [HPO] and [LPO].