5-8
5.14.3 Restore Recent Settings Automatically
When the ventilator is used on the same patient after powered on, the system adopts recent
settings automatically.
NOTE
Records the system saves automatically include reference loop, monitored trend,
event log (including alarm log), setup trend, special function measured values
(including PEEPi, NIF, P0.1, and P-V Tool measured values), patient setup and
equipment setup (including alarm setup). When there are changes in these data,
the system stores the changed data in the flash memory chips of the main board
automatically. When the ventilator restarts, the data are restored automatically.
5.15 Transfer Settings
You can export or import settings, while unit is in standby.
To export settings,
1. Insert the USB memory into the USB connector of the ventilator.
2. Select [Setup]→[Maintain]→[User]→enter the required password→[Data Transfer].
3. Select [Export Settings] to save the ventilator’s current settings and default settings to
the USB memory.
To import settings,
1. Insert the USB memory into the USB connector of the ventilator.
2. Select [Setup]→[Maintain]→[User]→enter the required password→[Data Transfer].
3. Select [Import Settings] to load the settings in the USB memory to the ventilator.